Skip to content

Public FDA record

Z-2753-2026: FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac

Recall record Z-2753-2026 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2753-2026
Company in FDA record
Boston Scientific Corporation
FDA classification date
Jul 17, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Boston Scientific is replacing specific FARASTAR PFA generators which have a configuration that is not compatible for use with the FARAPOINT PFA catheter.

Source facts

Action
Boston Scientific issued an Urgent Field Safety Notice - Urgent Medical Device Recall notice to its consignees out side the US on 6/11/2026 via hand delivered letter. The notice explained the issue and requested the following actions be taken: Instructions: 2- Allow Boston Scientific representative(s) to replace affected FARASTAR PFA generators. 3- Please complete the attached Verification Form even if you do not have any product to return and send it to your local Boston Scientific office 4- If you have products to return, please package them in an appropriate shipping box. After receipt of the Verification Form, Boston Scientific will contact you to arrange return. 6- Please pass this notice to any health professional from your organization that needs to be aware, and to any organization where the potentially affected devices have been transferred (if appropriate). Please provide Boston Scientific with details of any affected devices that have been transferred to other organizations (if appropriate). The National Competent Authority has been informed of this communication.
Initiated
2026-06-11
Root cause
Under Investigation by firm
Distribution
International distribution in the countries of Italy, Austria, France, Germany, Belgium.
Recall number
Z-2753-2026
Recall status
Open, Classified
CDRH record ID
221149
FDA classified
2026-07-17
Recall event number
99251

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗