Open original FDA record ↗Recall record
- FDA identifier
- Z-2755-2017
- Company in FDA record
- Medtronic Navigation, Inc.
- FDA classification date
- Jul 3, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.
Source facts
- Action
- Medtronic sent an Urgent Medical Device Safety Alert dated June 1, 2017, to all affected customers. Actions to be taken: 1. Please examine your inventory and if any of the affected products listed above are found, immediately quarantine them for return to Medtronic. 2. Complete and follow instruction on the attached consignee response form. Sign and date the bottom of the form and then return the form to Medtronic at RS.NavFCAmedtronic.com or fax to Medtronic Technical Services at 651-367-7075. If you have affected products, contact Medtronic Technical Services at 1-800 595-9709 to receive a return material authorization (RMA) and arrange for their return and no charge replacements. Once an RMA number is obtained, ship the affected product to Medtronic. For further questions please call (720) 890-3241.
- Initiated
- 2017-06-07
- Root cause
- Component design/selection
- Terminated
- 2018-06-29
- Distribution
- US Distribution to the states of : CA, GA, MN, FL, NY, IL, TX
- Recall number
- Z-2755-2017
- Recall status
- Terminated
- CDRH record ID
- 156887
- FDA classified
- 2017-07-03
- Recall event number
- 77623