Z-2755-2019: APS Accelerator modular systems, 208-230 V-, 50/60 Hz, 1800 VA, IVD, RX, Class 2 Laser Product,

Recall record Z-2755-2019 for Abbott Laboratories, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2755-2019
Company in FDA record
Abbott Laboratories, Inc
FDA record creation date
Sep 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm identified customer sites where modules which the safety pacemaker label is missing. Information regarding the label included in the Operations Manual has been updated to specify: 1) the risk of malfunction applies to anyone with a pacemaker who may work on these modules; 2) the minimum safety distance has increased to 200 mm (7.87") from transport mechanisms. Any person with a pacemaker must not get closer than this distance from modules axes.

Source facts

Action
On September 7, 2018, Abbott sent all affected consignees a "Product Correction Urgent - Immediate Action required" letter along with an "Attachment A" containing serial numbers. The letter informed consignees of the following: 1.Please be aware that personnel fitted with a pacemaker/implanted heart defibrillator must not handle or work on these modules at distances less than 200 mm even if the warning label is missing. 2. Your Abbott representative will be contacting you to schedule time to place a new hazard label on the APS modules. In the interim, please remove the last page of this letter and affix a copy of the label on the cover of these modules until the official pacemaker label is available. 3.Please review the attached Errata Sheet and save it with the Operations Manual for future reference regarding the safety distance required for pacemakers/implanted heart defibrillator. 4. Please retain this letter for your laboratory records. 5. If you or any of the health care providers you serve have any questions regarding this information, U.S. Customers please contact Customer Service at 1877 4ABBOTT (available 24 hours a day, 7 days a week). 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online at (http://www.fda.gov/MedWatch/report.htm), by mail (http://www.fda.gov/MedWatch/getforms.htm), by phone (18003321088), or by fax (1800FDA 0178).
Initiated
2018-09-07
Root cause
Component change control
Terminated
2021-02-02
Distribution
US: TX, VA, AZ, NV, MA, LA, MN, PA, SC, NJ, NY, CA, IL, IN, WA, FL, TN, PA, KS, ND, NC, CT OUS: Argentina, Australia, Belgium, Canada, China, Denmark, France, Greece, Hong Kong, Ireland, Italy, New Zealand, Norway, Oman, Pakistan, Philippines, Portugal, Russia, Saudi Arabia, Singapore, Spain, Taiwan, Turkey, UAE, UK, Venezuela
Recall number
Z-2755-2019
Recall status
Terminated
CDRH record ID
176042
FDA record created
2019-09-26
Recall event number
83651

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗