Skip to content

Public FDA record

Z-2755-2020: SIEMENS Healthineers Luminos dRF Max Model: 10762471- Product Usage: is a device intended to visualize a

Recall record Z-2755-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2755-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Aug 5, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis

Source facts

Action
Siemens Medical Solutions USA, Inc. has initiated a Customer Safety Advisory Notice to all affected customers via XP028/19/S January 2020. Additionally, a solution to eliminate the root cause of this problem is being distributed via XP008/20/S March 2020. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. New corrective software VE10Y became available June 22, 2020, and it is being distributed via an Update Instruction XP008/20/S. This update fixes the initial reported issue and the sporadic image loss issue noted by customers who had already installed VE10W (XP029/19/S). The original Customer Safety Advisory Notice (XP028/19/S) and the information letter distributed in March 2020 remain applicable. This resolution will be provided free of charge. Your local Siemens Healthineers Service organization will contact you to arrange a date to perform this corrective action . Questions contact the service organization for an earlier appointment at 1-800-888-7436
Initiated
2020-01-08
Root cause
Device Design
Terminated
2022-06-21
Distribution
US Nationwide distribution.
Recall number
Z-2755-2020
Recall status
Terminated
CDRH record ID
182399
FDA record created
2020-08-05
Recall event number
85999

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗