Open original FDA record ↗Recall record
- FDA identifier
- Z-2756-2016
- Company in FDA record
- Sorin Group USA, Inc.
- FDA classification date
- Sep 8, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Cannula With Ballon because the tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.
Source facts
- Action
- LivaNova issued an Urgent Safety Alert letter via certified mail on August 12, 2016 to all affected customers. The letter instructed customers to check their inventory to determine if the Cannulae belongs to the product code and lots listed. If the customer does not have any of the products described in the communication, please complete the response form below and return it per the instructions indicated on the form; otherwise, please check the tip angle to determine if it is conforming to specifications. If specification is not met, please file a complaint through your normal process whether you decide to keep it and use it or return the product to Sorin Group; if you are not willing to use it, return the affected parts to Sorin Group USA. Please contact Customer Service 1-S00-650-2623 or email CustomerService@livanova.com to have a Return Material Authorization (RMA) issued. Complete the Customer Response Form attached to this letter to acknowledge receipt of this letter. Please return this form no later than September 15, 2016. For questions regarding this recall call 303-467-6306.
- Initiated
- 2016-08-12
- Root cause
- Process control
- Terminated
- 2017-03-27
- Distribution
- Worldwide Distribution - US including IL, MO, TX. Internationally to Canada and Italy.
- Recall number
- Z-2756-2016
- Recall status
- Terminated
- CDRH record ID
- 148850
- FDA classified
- 2016-09-08
- Recall event number
- 74934