Open original FDA record ↗Recall record
- FDA identifier
- Z-2756-2019
- Company in FDA record
- Medline Industries Inc
- FDA record creation date
- Sep 26, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The recall is being conducted due to a potential defective component on the circuit board causing the charger base to overheat which could result in a charger base malfunction.
Source facts
- Action
- Medline Industries initiated the recall on 09/06/2019 by letter. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Immediately check your stock for the affected item number and lot number listed on the destruction form. Quarantine all affected product. Destroy all product belonging to the affected lot numbers listed on the destruction form. 2. Please return the completed enclosed destruction form listing the quantity of affected product destroyed. Even if you do not have any affected product please complete and return the form. For questions contact 866-359-1704.
- Initiated
- 2019-09-06
- Root cause
- Process control
- Distribution
- US Nationwide Distribution
- Recall number
- Z-2756-2019
- Recall status
- Open, Classified
- CDRH record ID
- 176259
- FDA record created
- 2019-09-26
- Recall event number
- 83740