Z-2759-2026: Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and delive

Recall record Z-2759-2026 for ICU Medical Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2759-2026
Company in FDA record
ICU Medical Inc.
FDA classification date
Jul 17, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTION notification letter dated 6/24/26 was sent to customers. Actions for Users: When using the device, all instructions, including warnings and cautions contained in the Instructions for Use must be followed with heightened awareness. Please complete the following actions below. 1. Locate potentially affected devices from the Lots listed in Table 1 above. 2. Destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 3. Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used at another location, please ensure that this communication is delivered there. 4. Complete and return the attached Customer Response Form to MarketAction@mailac.custhelp.com within 10 days of receipt to acknowledge your understanding of this notification. Required Actions for Distributors: When using the device, all instructions, including warnings and cautions contained in the Instructions for Use must be followed with heightened awareness. Please complete the following actions below. 1. Locate potentially affected devices from the Lots listed in Table 1 above. 2. Destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 3. Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used at another location, please ensure that this communication is delivered there. 4. Complete and return the attached Customer Response Form to MarketAction@mailac.custhelp.com within 10 days of receipt to acknowledge your understanding of this notification. 5. DISTRIBUTORS: Immediately forward this notice to all users to whom you distributed these products. IC
Initiated
2026-06-24
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-2759-2026
Recall status
Open, Classified
CDRH record ID
221197
FDA classified
2026-07-17
Recall event number
99289

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗