Skip to content

Public FDA record

Z-2760-2017: Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluo

Recall record Z-2760-2017 for Olympus Scientific Solutions Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2760-2017
Company in FDA record
Olympus Scientific Solutions Americas
FDA record creation date
Aug 2, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Instruments may, under certain specific circumstances listed below, fail to stop when a test is initiated and no sample is present (proximity shut down fails to end test).

Source facts

Action
Olympus Scientific Solutions Americas Corp will bring defect into compliance: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Tracking of customer/owner self-installation(s) of the software update should be done to the maximum possible extent via serial-number-tracked downloads, follow-up email reporting, return- mail card or by other similarly effective means. For further questions, please call (781) 419-3900.
Initiated
2017-07-11
Root cause
Radiation Control for Health and Safety Act
Terminated
2023-05-03
Distribution
Worldwide Distribution - USA (nationwide) Distribution
Recall number
Z-2760-2017
Recall status
Terminated
CDRH record ID
157835
FDA record created
2017-08-02
Recall event number
77759

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗