Z-2760-2026: Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professiona

Recall record Z-2760-2026 for Ferno-Washington Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2760-2026
Company in FDA record
Ferno-Washington Inc
FDA classification date
Jul 17, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.

Source facts

Action
On June 19. 2026, the firm began notifying customers via Urgent Medical Device Correction letters. Customers were instructed to contact Ferno for assistance if abnormal rail movement, increased resistance during rail extension or retraction, or other unusual rail operation is observed. Ferno will contact all affected customers directly and dispatch trained field service personnel to perform the correction onsite at customer locations. The correction does not require removal of the PF2 Fastening System from service prior to the scheduled repair. The field correction consists of replacement of the affected roller assembly with an updated assembly manufactured in accordance with revised assembly requirements intended to eliminate the potential for interference and subsequent roller deformation.
Initiated
2026-06-19
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, GA, IA, ID, IL, IN, KY, LA, MD, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Poland, Netherlands, Canada, Germany, Norway, United Kingdom, Denmark and New Zealand.
Recall number
Z-2760-2026
Recall status
Open, Classified
CDRH record ID
221198
FDA classified
2026-07-17
Recall event number
99291

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗