Open original FDA record ↗Recall record
- FDA identifier
- Z-2761-2026
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Jul 20, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities.
Source facts
- Action
- On July 7, 2026, the firm issued "Urgent: Medical Device Correction" letters to consignees. Customers were informed of the product issue. While the identified 4K UHD LCD monitor may be used with VISERA S, VISERA ELITE III, EVIS X1, and EU-ME3 systems, the firm notes that the affected HDMI or DP connectors are more commonly used with the VISERA S video system center. Its use with the three other systems does not typically involve the affected HDMI or DP connectors. Customers are requested to examine their equipment, looking for the identified monitor model number and serial numbers listed in the customer letter. They should then acknowledge receipt of the letter through the firm's recall web portal. The firm will coordinate repair of the monitors. If customer needs to use the monitor prior to its repair, the firm recommends that a spare unit be available in case issues are encountered with the primary unit during use.
- Initiated
- 2026-07-07
- Root cause
- Nonconforming Material/Component
- Distribution
- Worldwide - US Nationwide distribution in the states of OH, CA, NJ, FL and the country of Japan.
- Recall number
- Z-2761-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221196
- FDA classified
- 2026-07-20
- Recall event number
- 99288