For systems upgraded to syngo Dynamics version 9.0, if a legacy report generated by version 3.x or earlier is opened in the version 9.0 portal, the legacy report may be deleted and replaced with a new report from the version 9.0 reporting system or a blank report.
Source facts
Action
Siemens Medical Solutions, USA, Inc sent a letter entitled "Customer Safety Advisory Notice" to all direct accounts on May 10, 2011. The letter stated the name and version of the software, problem noted, a work-around to be implemented until a service patch is created to fix the problem, and that the letter should be placed in the software's operator's manual to be seen by all who utilize the software.
Initiated
2011-05-10
Root cause
Software design
Terminated
2012-07-09
Distribution
Nationwide Distribution
Recall number
Z-2762-2011
Recall status
Terminated
CDRH record ID
100383
FDA classified
2011-07-08
Recall event number
58830