Skip to content

Public FDA record

Z-2762-2011: syngo Dynamics Picture Archiving Communication System The software is a picture archiving and communicat

Recall record Z-2762-2011 for Siemens Medical Solutions, USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2762-2011
Company in FDA record
Siemens Medical Solutions, USA, Inc
FDA classification date
Jul 8, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

For systems upgraded to syngo Dynamics version 9.0, if a legacy report generated by version 3.x or earlier is opened in the version 9.0 portal, the legacy report may be deleted and replaced with a new report from the version 9.0 reporting system or a blank report.

Source facts

Action
Siemens Medical Solutions, USA, Inc sent a letter entitled "Customer Safety Advisory Notice" to all direct accounts on May 10, 2011. The letter stated the name and version of the software, problem noted, a work-around to be implemented until a service patch is created to fix the problem, and that the letter should be placed in the software's operator's manual to be seen by all who utilize the software.
Initiated
2011-05-10
Root cause
Software design
Terminated
2012-07-09
Distribution
Nationwide Distribution
Recall number
Z-2762-2011
Recall status
Terminated
CDRH record ID
100383
FDA classified
2011-07-08
Recall event number
58830

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗