- FDA identifier
- Z-2762-2015
- Company in FDA record
- Revmedx Inc
- FDA classification date
- Sep 11, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The XGAUZE Trauma Dressing was recalled because the packages may acquire small tears or pinholes during shipping due to use of shipping containers that lacked adequate protective materials. If present, these small tears or pinholes may disrupt package integrity and render the device unsterile.
Source facts
- Action
- RevMedx, Inc. sent the Urgent Medical Device Correction and Removal XGAUZE Trauma Dressing letter to US Customers on August 12, 2015. Customers are instructed to inspect packages for signs of damages or loss of vacuum. The affected packages will appear puffy and the dressing will be loose inside. The non-affected packages will have a good vacuum intact seal and the packaging material should contour snugly against the dressing. Customers are advised to return any packages appear to have damaged or loss of vacuum. Any undamaged packages may be used as indicated. Customers with question can call the firm at (503) 218-2172 and fax the Return Response Form via fax to (503) 218-2274 or send via email to info@revmedx.com.
- Initiated
- 2015-08-12
- Root cause
- Package design/selection
- Terminated
- 2015-09-17
- Distribution
- US Distribution to the states of : Arizona, Idaho, North Carolina, Tennessee and Virginia.
- Recall number
- Z-2762-2015
- Recall status
- Terminated
- CDRH record ID
- 139586
- FDA classified
- 2015-09-11
- Recall event number
- 72019