Open original FDA record ↗Recall record
- FDA identifier
- Z-2762-2017
- Company in FDA record
- Fisher Diagnostics
- FDA record creation date
- Jul 25, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list 100651, lot 890199. While the kit performs accurately at FDP concentrations greater than 20 g/ml, there is an observed reduction in sensitivity when FDP levels are less than 20 g/ml. This reduction in sensitivity has been tied to FDP Latex Reagent lot 890199 and may result in false negatives.
Source facts
- Action
- ThermoFisher Scientific sent an MEDICAL DEVICE RECALL Letter dated May 16, 2017, to all affected customers and response forms were sent to customers on May 19, 2017. Customers are informed of the recall and potential health risks and instructed to review their test results. Review your inventory and any product on hand is to be discarded. --- The firm ceased distribution of the lots and began its investigation. The held product will ultimately be destroyed. Level A effectiveness checks will be conducted to the consignees and repeated attempts will be made to non-responding customers. The firm will track all mailed, emailed and faxed return forms for the duration of the recall. For further questions, please call (540) 869-8188.
- Initiated
- 2017-05-19
- Root cause
- Under Investigation by firm
- Terminated
- 2018-06-29
- Distribution
- USA (nationwide) including Puerto Rico, and Internationally to Colombia
- Recall number
- Z-2762-2017
- Recall status
- Terminated
- CDRH record ID
- 155789
- FDA record created
- 2017-07-25
- Recall event number
- 77340