Skip to content

Public FDA record

Z-2762-2017: Pacific Hemostasis FDP (Fibrinogen Degradation Products) Assay Kit (30 Determinations), REF/Model 100650,

Recall record Z-2762-2017 for Fisher Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2762-2017
Company in FDA record
Fisher Diagnostics
FDA record creation date
Jul 25, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list 100651, lot 890199. While the kit performs accurately at FDP concentrations greater than 20 g/ml, there is an observed reduction in sensitivity when FDP levels are less than 20 g/ml. This reduction in sensitivity has been tied to FDP Latex Reagent lot 890199 and may result in false negatives.

Source facts

Action
ThermoFisher Scientific sent an MEDICAL DEVICE RECALL Letter dated May 16, 2017, to all affected customers and response forms were sent to customers on May 19, 2017. Customers are informed of the recall and potential health risks and instructed to review their test results. Review your inventory and any product on hand is to be discarded. --- The firm ceased distribution of the lots and began its investigation. The held product will ultimately be destroyed. Level A effectiveness checks will be conducted to the consignees and repeated attempts will be made to non-responding customers. The firm will track all mailed, emailed and faxed return forms for the duration of the recall. For further questions, please call (540) 869-8188.
Initiated
2017-05-19
Root cause
Under Investigation by firm
Terminated
2018-06-29
Distribution
USA (nationwide) including Puerto Rico, and Internationally to Colombia
Recall number
Z-2762-2017
Recall status
Terminated
CDRH record ID
155789
FDA record created
2017-07-25
Recall event number
77340

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗