Blackrock Microsystems, LLC announces a voluntary field action for the Blackrock NeuroPort Biopotential Signal Processing System because the Neural Signal Amplifier and the Patient Cable are incorrectly labeled as Type CF Applied Parts.
Source facts
Action
Blackrock Microsystems sent an Urgent Medical Device letter dated August 12, 2016 via email. The letter identified the affected product, problem and actions to be taken. Customers were instructed not use the amplifier in application requiring a type CF applied part; perform a physical count of the inventory of amplifiers and patient cables with the part numbers listed. For questions regarding this Correction call 801-582- 5533, ext. 240.
Initiated
2016-08-10
Root cause
Labeling False and Misleading
Terminated
2017-08-21
Distribution
Worldwide Distribution - US Nationwide and Canada,
Recall number
Z-2763-2016
Recall status
Terminated
CDRH record ID
148705
FDA classified
2016-09-08
Recall event number
74893