Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).
Source facts
Action
An Urgent Medical Device Recall Letter was sent during the week of 9/17/2018 to all affected customers via mail and or email. Customers were instructed to discontinue use and discard remaining product. Customers were also instructed to consider a review of the Cortisol results generated with the affected lot number at the discretion of the Medical Director.
Initiated
2018-09-18
Root cause
Process design
Terminated
2024-02-21
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Lithuania, Mexico, Panama, Serbia, South Africa, and Vietnam.
Recall number
Z-2763-2019
Recall status
Terminated
CDRH record ID
173881
FDA record created
2019-09-30
Recall event number
83163