Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
Source facts
Action
In a Quality Notification dated December 19, 2018, the firm informed its distribution centers of the issue. Contact Centers were provided with updated troubleshooting guides for Aviva II and Performa II meters in August 2018 with instructions for phone solving power issues for meters produced between June 2016 and June 2017.
The firm has taken corrective actions to address the issue.
Customers should contact Accu-Chek Customer Care line at 1-800-358-4866, Monday through Friday, between the hours of 8:00 am and 8:00 pm Eastern Standard Time
Initiated
2017-06-12
Root cause
Process change control
Terminated
2020-10-30
Distribution
Worldwide distribution - US Nationwide distribution.
Recall number
Z-2765-2019
Recall status
Terminated
CDRH record ID
175782
FDA classified
2019-09-30
Recall event number
83574