Skip to content

Public FDA record

Z-2766-2019: Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System

Recall record Z-2766-2019 for Roche Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2766-2019
Company in FDA record
Roche Diabetes Care, Inc.
FDA classification date
Sep 30, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).

Source facts

Action
In a Quality Notification dated December 19, 2018, the firm informed its distribution centers of the issue. Contact Centers were provided with updated troubleshooting guides for Aviva II and Performa II meters in August 2018 with instructions for phone solving power issues for meters produced between June 2016 and June 2017. The firm has taken corrective actions to address the issue. Customers should contact Accu-Chek Customer Care line at 1-800-358-4866, Monday through Friday, between the hours of 8:00 am and 8:00 pm Eastern Standard Time
Initiated
2017-06-12
Root cause
Process change control
Terminated
2020-10-30
Distribution
Worldwide distribution - US Nationwide distribution.
Recall number
Z-2766-2019
Recall status
Terminated
CDRH record ID
175862
FDA classified
2019-09-30
Recall event number
83574

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗