Open original FDA record ↗Recall record
- FDA identifier
- Z-2767-2015
- Company in FDA record
- Nico Corp.
- FDA classification date
- Sep 15, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.
Source facts
- Action
- The firm, Nico Corporation, sent an "URGENT: NICO Console Recall" notification dated 7/31/15 via fax to its customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The customers were instructed to return the console and complete and return the Recall Response Form via fax to: 317-683-0305 or email to: Jay.Dittman@niconeuro.com. The firm will provide customers with a Myriad Console loaner kit. If you have any questions, contact Director of Quality and Regulatory, at 317-660-7118 ext 104 or email to: jay.dittman@nicneuro.com.
- Initiated
- 2015-07-29
- Root cause
- Process control
- Terminated
- 2015-09-22
- Distribution
- US Distribution to states of: MD, IN, CA, and NC.
- Recall number
- Z-2767-2015
- Recall status
- Terminated
- CDRH record ID
- 139582
- FDA classified
- 2015-09-15
- Recall event number
- 72015