Novadaq is informing SK6000 PAQ users that a pinhole through the drape pouch film material was identified in some SK6000 PAQs during shipping validation testing.
Source facts
Action
The firm, Novadaq, sent an "URGENT DEVICE CORRECTION: SK6000 PAQ Drape Pouch" letter dated 8/18/2016 to affected consignees. The Firm instructed consignee to cease distribution of affected product; accept the shipping-validated replacement PAQ; complete and sign the enclosed "Acknowledgment of Receipt" form, scan it and email to quality@novadaq.com.
If you have any questions, contact Director of Regulatory Affairs at 1-905-629-3822 ext 205 or email: jpendlebury@novadaq.com.
.
Initiated
2016-08-18
Root cause
Packaging
Terminated
2016-12-09
Distribution
US Distribution in states of: FL, MI, and VA.
Recall number
Z-2767-2016
Recall status
Terminated
CDRH record ID
149092
FDA record created
2016-09-08
Recall event number
75006