Skip to content

Public FDA record

Z-2767-2019: Aquilex Fluid Control System component: Bag deflector, REF AQL-100CBS Product Usage: Distends uterus w

Recall record Z-2767-2019 for WOM World of Medicine AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2767-2019
Company in FDA record
WOM World of Medicine AG
FDA record creation date
Sep 30, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The deficit displayed by the pump can differ from the real deficit, leading potentially to patient risk.

Source facts

Action
On August 14, 2019, W.O.M. World of Medicine GmbH issued a Medical Device Recall notice to their customer (distributor). On September 5, 2019, Hologic issued Medical Device Recall notices to their customers via USPS priority mail with delivery confirmation. Customers were advised to take the following actions: 1) Perform a visual check of your Aquilex system looking for obvious damage to the bag deflector using the criteria identified within the customer notice. 2) If you do not notice any obvious damage using the criteria listed within the customer letter, and your unit passes the below mentioned Fluid Deficit Test you may continue to use your unit until you receive your replacement. 3) If you notice any damage or if any of the mentioned criteria are not met please stop using your Aquilex Fluid Control system. Customers with questions, please contact Hologic Technical Support at 1-855-898-5025. A Hologic representative will assist you with your return, as well as coordinate a replacement unit for you.
Initiated
2019-08-14
Root cause
Nonconforming Material/Component
Terminated
2020-05-20
Distribution
US Nationwide Distribution
Recall number
Z-2767-2019
Recall status
Terminated
CDRH record ID
176196
FDA record created
2019-09-30
Recall event number
83712

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗