Open original FDA record ↗Recall record
- FDA identifier
- Z-2768-2015
- Company in FDA record
- MedtestDx, Inc.
- FDA classification date
- Sep 15, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Vial to vial variation in the fill volume that could affect control recovery issues and cause a control result to fall outside the suggested recovery ranges. An under-fill of the vials was found ranging as low as 4.63 ml. When used to reconstitute HNC/HEC, control values were shown to be out of range running on the high end.
Source facts
- Action
- On 5/15/2015, MedTest DX Field Correction Notice notifications were sent to the affected customers via electronic mail and US mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Any questions concerning the corrective action may be directed to MedTest Technical Service Department at 1-800-757-5313.
- Initiated
- 2015-05-15
- Root cause
- Process design
- Terminated
- 2016-05-23
- Distribution
- US Nationwide Distribution.
- Recall number
- Z-2768-2015
- Recall status
- Terminated
- CDRH record ID
- 139557
- FDA classified
- 2015-09-15
- Recall event number
- 71993