Skip to content

Public FDA record

Z-2768-2020: Beacon EUS Access System Preloaded Access 90degree Needle sterile, single-use, endoscopic ultrasound dev

Recall record Z-2768-2020 for Covidien Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2768-2020
Company in FDA record
Covidien Llc
FDA classification date
Aug 5, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.

Source facts

Action
Medtronic sent a letter notification to customers dated July 06, 2020 recalling effected lots of product. This letter instructed customers to discontinue use of the product and quarantine effected inventory. Customers were also instructed to return effected product and complete recall response forms accordingly.
Initiated
2020-06-26
Root cause
Under Investigation by firm
Distribution
world wide distribution
Recall number
Z-2768-2020
Recall status
Open, Classified
CDRH record ID
182412
FDA classified
2020-08-05
Recall event number
86001

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗