Skip to content

Public FDA record

Z-2768-2026: Ventriculostomy Kit, REF: 46154

Recall record Z-2768-2026 for Medtronic Neurosurgery. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2768-2026
Company in FDA record
Medtronic Neurosurgery
FDA classification date
Jul 21, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ventriculostomy devices, used during ventriculostomy procedures to gain access to the ventricular system, may exhibit endotoxin levels that exceed the acceptable limit for devices that come into contact with cerebrospinal fluid; endotoxin exposure can cause an acute inflammatory process.

Source facts

Action
On 6/17/2026, recall notices were mailed and emailed to customers asking them to do the following: 1) Identify, quarantine, and return affected devices to the firm. 2) Complete and return the Customer Confirmation Form via email to neuro.quality@medtronic.com 3) This notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. For additional information, email the firm at rs.mnsprodexperiencehelp@medtronic.com
Initiated
2026-06-17
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of UT, AL, CA, NC, GA, AR, LA, KY, LA, KY, IL, OK and the country of Costa Rica.
Recall number
Z-2768-2026
Recall status
Open, Classified
CDRH record ID
221365
FDA classified
2026-07-21
Recall event number
99342

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.