Open original FDA record ↗Recall record
- FDA identifier
- Z-2771-2020
- Company in FDA record
- Ortho Clinical Diagnostics
- FDA record creation date
- Aug 7, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The reagent bottles may be swapped within the reagent pack, which may lead to false negative results or delayed results due to calibration failures, or quality control failures.
Source facts
- Action
- On July 17, 2020, the firm issued Urgent Product Correction Notification letters to affected customers. Customers were informed of the potential for erroneous patient sample results, calibration errors, and QC failures when using the impacted product. Customers were asked to do the following; " Visually inspect affected VITROS Anti-SARS-CoV-2 Total (Lots 0010-0037) and IgG (Lots 0010-0013) Reagent Packs to ensure correct reagent bottle configuration " Set aside any pack in which the reagent bottles are positioned incorrectly, so the lot number and pack id can be provided to Ortho. Complete and submit the enclosed Request for Credit Form, including the pack information, to receive credit for an affected pack. " Complete the Confirmation of Receipt form no later than July 28, 2020. " Post this notification by each system that processes an affected lot. " Please forward this notification if the product was distributed outside of your facility. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
- Initiated
- 2020-07-17
- Root cause
- Process control
- Terminated
- 2021-07-16
- Distribution
- Worldwide distribution - US Nationwide distribution.
- Recall number
- Z-2771-2020
- Recall status
- Terminated
- CDRH record ID
- 182548
- FDA record created
- 2020-08-07
- Recall event number
- 86038