Open original FDA record ↗Recall record
- FDA identifier
- Z-2771-2026
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Jul 22, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
Source facts
- Action
- Boston Scientific initially notified their consignees via teleconference on 6/19/2026 and followed with a notice distributed on the same date via email. The notice explained the issue, potential harms, and requested the following: "Required User Actions " Follow the instructions provided in the TheraSphere 360 Ordering Quick Guide. " Remove and reselect vials if inputs into the Activity calculator (Lung Shunt Fraction, Anticipated Residual Waste, and/or Target Volume) are changed after any vials were added to the cart. " Verify that the vial information displayed in the cart and treatment documentation match the final calculation inputs. " Review order information in the attachment to ensure treatment and documentation are as intended. Contact your TheraSphere Consultant to address any impacted treatments or with any questions."
- Initiated
- 2026-06-19
- Root cause
- Device Design
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.
- Recall number
- Z-2771-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221119
- FDA classified
- 2026-07-22
- Recall event number
- 99236