Z-2771-2026: Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image

Recall record Z-2771-2026 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2771-2026
Company in FDA record
Boston Scientific Corporation
FDA classification date
Jul 22, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.

Source facts

Action
Boston Scientific initially notified their consignees via teleconference on 6/19/2026 and followed with a notice distributed on the same date via email. The notice explained the issue, potential harms, and requested the following: "Required User Actions " Follow the instructions provided in the TheraSphere 360 Ordering Quick Guide. " Remove and reselect vials if inputs into the Activity calculator (Lung Shunt Fraction, Anticipated Residual Waste, and/or Target Volume) are changed after any vials were added to the cart. " Verify that the vial information displayed in the cart and treatment documentation match the final calculation inputs. " Review order information in the attachment to ensure treatment and documentation are as intended. Contact your TheraSphere Consultant to address any impacted treatments or with any questions."
Initiated
2026-06-19
Root cause
Device Design
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.
Recall number
Z-2771-2026
Recall status
Open, Classified
CDRH record ID
221119
FDA classified
2026-07-22
Recall event number
99236

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗