Z-2772-2020: OmniTom Computed tomography x-ray system, Model NL5000, all NL5000 systems with software version before 0

Recall record Z-2772-2020 for NeuroLogica Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2772-2020
Company in FDA record
NeuroLogica Corporation
FDA record creation date
Aug 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There are potential issues related to the operational state of the scanner, specifically for transport and inadvertent system shutdowns. There could be a possibility of potential delay in patient treatment if proper use instructions are not followed.

Source facts

Action
Urgent Medical Device Correction notification letters were sent to customers beginning 7/15/20. Users may continue to use the NL5000 system if proper instructions for use are followed as described in user manual (1-NL5000-060). This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please transfer this notice to other organizations on which this action has an impact. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. The system label is located near the main input plug or on the side of the system. NeuroLogica will undertake the voluntary corrective action considering the possibility of delay in patient treatment if the scanner is not operational. The hardware and software upgrades provide additional safeguards by improving the design to minimize any risks of the above situations from happening. Service engineers will correct all affected devices free of charge and contact you to arrange for the correction. All corrections in the field will be completed by December 31, 2020. If you need any further information or have any concerns with this issue, then please contact your local representative. Customer Reply Form This reply form is to confirm receipt of the enclosed NeuroLogica Medical Device Field Safety Notice [FSN-NL5000_071620] regarding the potential issues related to the operational state of the scanner and the required software and hardware upgrades. Please read the Notice and indicate the appropriate answers to the questions below.
Initiated
2020-07-15
Root cause
Device Design
Terminated
2021-07-13
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, India, Portugal, Singapore, Slovak Republic, Thailand, Turkey, UK.
Recall number
Z-2772-2020
Recall status
Terminated
CDRH record ID
182577
FDA record created
2020-08-07
Recall event number
86052

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗