Skip to content

Public FDA record

Z-2772-2026: Halyard STANDARD LINE INSERTION PACK. Model/Catalog Number: PVHM234-01. Convenience kit.

Recall record Z-2772-2026 for AVID Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2772-2026
Company in FDA record
AVID Medical, Inc.
FDA classification date
Jul 22, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Defects in the silicone seal, which may result in "doming" of the silicone seal due to improper sealing, potentially leading to fluid occlusion or leakage.

Source facts

Action
AVID Medical notified consignees on about 05/19/2026 via email. Consignees were instructed to identify, segregate, and quarantine all impacted kits, and add the provided warning labels to affected AVID kits to ensure all users are aware of discarding the impacted component prior to use. Distributors were instructed to ensure all end-users are properly notified if affected kits have been further distributed. Customers/End-users were instructed that if they have received product from a distributor, to contact the distributor or your sales representative with any questions. Recipients were requested to complete and return the provided Recall Response Form.
Initiated
2026-05-19
Root cause
Process change control
Distribution
U.S. Nationwide distribution in the state of California.
Recall number
Z-2772-2026
Recall status
Open, Classified
CDRH record ID
221147
FDA classified
2026-07-22
Recall event number
99250

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗