Open original FDA record ↗Recall record
- FDA identifier
- Z-2773-2015
- Company in FDA record
- Pointe Scientific, Inc.
- FDA classification date
- Sep 16, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported as the control will be out of specifications.
Source facts
- Action
- On 12/12/2014, Field Correction Notice notifications were sent to the affected distributors via electronic mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Any questions concerning the corrective action may be directed to Pointe Scientifics Technical Service Department at 1-800-757-5313.
- Initiated
- 2014-12-12
- Root cause
- Process control
- Terminated
- 2015-10-29
- Distribution
- Worldwide distribution. US nationwide, Singapore, Kenya, Bolivia, Mexico, Greece, Guyana, and Philippines.
- Recall number
- Z-2773-2015
- Recall status
- Terminated
- CDRH record ID
- 139476
- FDA classified
- 2015-09-16
- Recall event number
- 71974