Open original FDA record ↗Recall record
- FDA identifier
- Z-2775-2016
- Company in FDA record
- Thermo Finnigan LLC
- FDA classification date
- Sep 9, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Thermo Fisher has determined that the Endura MD mass spectrometer instrument control software versions 1.0 and 1.0 SP! have a software defect which affect data accuracy.
Source facts
- Action
- ThermoFisher Scientific sent an Urgent Medical Device Field Correction letter dated July 19, 2016, to all affected customers via FedEx. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to discontinue use of the affected product until a company authorized Field Service Engineer installs corrected software to the instrument system. Customers were also instructed to communicate to all who operate the instrument and to complete the provided Acknowledgment Form, sign, and return it via email FieldActionTeam@thermofisher.com. Customers with questions should contact the Field Action Team at FieldAction@thermofisher.com.
- Initiated
- 2016-07-19
- Root cause
- Software design
- Terminated
- 2017-02-21
- Distribution
- Worldwide distribution - US including FL, NY and Internationally to Japan.
- Recall number
- Z-2775-2016
- Recall status
- Terminated
- CDRH record ID
- 148564
- FDA classified
- 2016-09-09
- Recall event number
- 74837