Z-2775-2020: Verigene Processor SP with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP,

Recall record Z-2775-2020 for Luminex Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2775-2020
Company in FDA record
Luminex Corporation
FDA record creation date
Aug 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call.

Source facts

Action
On 06/10/2020, the firm sent a letter to customers titled, "MEDICAL DEVICE FIELD ACTION", regarding an instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call. The firm has developed a software solution that identifies and alerts users to the issue. The firm will contact customers to verify their address to ship the software material and provide installation information. Once the update has been completed, customers need to email a screen shot of the updated software, as well as a completed Notice of Destruction to Global Support Services at support@luminexcorp.com to complete the upgrade process. The firm is also stating that until the updated software is installed, it is important, and part of good laboratory practice, that the customer follow the specific guidelines indicated in the current package inserts. Specifically, "Not Detected" test results from a positive blood culture should be evaluated further. The sub-culturing of blood cultures is necessary, per the intended use section within the Package Inserts in the Kits. For any questions contact the firm's Global Support Service at: 1.877.785.2323 (US) 1.512.381.4397 (OUS/CANADA) 512.219.5114 support@luminexcorp.com www.luminexcorp.com
Initiated
2020-06-15
Root cause
Device Design
Terminated
2023-06-30
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CN, CT, DE, DC, FL,GA, ID, IN, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NB, NV, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN TX, UT, VT, VA, WA, WV, WY, and WI. The countries of Austria, Belgium, Bulgaria, Canada, Chile, China, France, Germany, Greece, Hong Kong, Italy, Japan, Republic of Korea, Kuwait, Macau, Mexico, Netherlands, New Zealand, Republic of Russia, Saudi Arabia, Singapore, Spain, Thailand, Turkey, United Arab Emirate, and United Kingdom.
Recall number
Z-2775-2020
Recall status
Terminated
CDRH record ID
181969
FDA record created
2020-08-10
Recall event number
85872

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗