Skip to content

Public FDA record

Z-2775-2026: SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.

Recall record Z-2775-2026 for KENT IMAGING, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2775-2026
Company in FDA record
KENT IMAGING, INC.
FDA classification date
Jul 22, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.

Source facts

Action
On 6/30/2026, correction and removal notices were emailed to customers who were asked to do the following: 1) Discontinue use of the affected unit(s) and remove them from clinical service. 2) Contact Kent Imaging at customerservice@kentimaging.com to confirm return shipping instructions. 3) Pending return, keep the device charged and avoid allowing the battery to fully deplete, to prevent the unit from becoming temporarily inoperable in the interim. 4) Within 5 business days of receipt of this notice, please confirm receipt and current status of the listed unit(s) by replying to this letter or contacting Hitalo Arume directly. Distributor was asked to: a) Identify the current location/holder of each of these 6 units; b) Coordinate scheduling of a brief remote session between Kent Imaging's technical support team and each unit's user; and c) Confirm completion of each update on the attached Device Reconciliation Form. d) For units that require physical return, quarantine and stop distribution. For units distributed, identify each affected customer, coordinate on-site access for collection, and arrange return shipment to Kent Imaging. Questions can be directed to hitaloa@kentimaging.com
Initiated
2026-06-30
Root cause
Software design
Distribution
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
Recall number
Z-2775-2026
Recall status
Open, Classified
CDRH record ID
221505
FDA classified
2026-07-22
Recall event number
99406

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗