Open original FDA record ↗Recall record
- FDA identifier
- Z-2776-2026
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Jul 22, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.
Source facts
- Action
- Olympus notified consignees on about 06/16/2026 via letter. Consignees were informed that they many continue to use their UHI-4 until the offset adjustment is performed. However, as noted in the Instructions for Use (IFU), prepare a spare UHI-4 unit as a backup to ensure uninterrupted operation can be completed without complication in the case of a malfunction. Actions to take included to carefully read the notification letter and ensure all personnel are completely knowledgeable on the content of the notification, an Olympus field service representative will contact you to schedule a mutually convenient time to perform the offset adjustment, acknowledge receipt of the notification through the Olympus recall web portal, and forward the notification to other users who may have the affected products if you have further distributed it.
- Initiated
- 2026-06-16
- Root cause
- Nonconforming Material/Component
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, El Salvador, Germany, Honduras, Hong Kong, India, Ireland, Japan, Korea, Malaysia, Mexico, Myanmar, New Zealand, Paraguay, Peru, Russia, Singapore, Sri Lanka, Taiwan, Thailand, United Kingdom, Uruguay, Vietnam.
- Recall number
- Z-2776-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220849
- FDA classified
- 2026-07-22
- Recall event number
- 99109