Open original FDA record ↗Recall record
- FDA identifier
- Z-2777-2016
- Company in FDA record
- Integra LifeSciences Corporation
- FDA record creation date
- Sep 9, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Integra's post QA release review of historical product release test results for Meshed IDRT products identified the Peel Strength test average result was incorrectly calculated for a single Lot (Lot 105A00324750).
Source facts
- Action
- Integra LifeSciences Inc. sent an urgent voluntary medical device recall letter/recall acknowledgement and return form dated March 11. 2016. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory for the affected product and immediately stop using and remove from service. Customers were asked to complete the attached form and return by email or fax as indicated on the form. Customers with questions were advised to contact Customer Service at 1-888-601-0203. For questions regarding this recall call 609-275-2700.
- Initiated
- 2016-03-11
- Root cause
- Under Investigation by firm
- Terminated
- 2016-09-13
- Distribution
- Nationwide Distribution to PA
- Recall number
- Z-2777-2016
- Recall status
- Terminated
- CDRH record ID
- 144759
- FDA record created
- 2016-09-09
- Recall event number
- 73689