Z-2777-2026: Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based aco

Recall record Z-2777-2026 for Emano Metrics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2777-2026
Company in FDA record
Emano Metrics, Inc.
FDA classification date
Jul 22, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.

Source facts

Action
On 06/10/2026, a "Medical Device Correction" Letter was emailed to customers informing them that it would found that Emano Flow software versions v2.2.0-v2.3.4 processed audio at a sample rate that did not match the sample rate expected by the flow prediction algorithms (model version v2.0.0), affecting results generated between 03/22/2026 to 05/29/2026. Customer were also informed that, the correction (updated software) was deployed on May 29, 2026. An updated customer notice was emailed on 06/23/2026. Instructions provided to Customer included: 1. Review the corrected results for patients with Emano Flow data from the affected period, prioritizing cases where the data informed a clinical decision. 2. If customers have previously downloaded patient reports generated between 03/22/2026 and 05/29/2026, replace them with the corrected versions available in the Emano Flow clinic portal. A spreadsheet with secure links to all corrected reports is attached. 3. Please reply to this email to confirm receipt of this notice.
Initiated
2026-05-29
Root cause
Software Design Change
Distribution
U.S. Nationwide distribution in the states of AR, AZ, CA, CO, FL. IA, ID, NC, NJ, PA, TX and WI.
Recall number
Z-2777-2026
Recall status
Open, Classified
CDRH record ID
221128
FDA classified
2026-07-22
Recall event number
99241

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗