Some drive stand bolts may fail to meet mechanical specifications.
Source facts
Action
Varian Medical Systems, Inc sent a "Urgent Medical Device Correction Urgent Field Safety Notice" letter dated May 13, 2011 to all affected customers.
The letter included a description of product, problem and recommendations. This includes performing daily QA to detect isocenter shift and the use of independent patient position system. Customers are advised to discontinue its use if a shift is detected and to contact Varian Medical Systems for a replacement stand bolts and adjustment if necessary.
For additional information please contact the Varian Oncology Help Desk at 888-VARIAN5 (888- 827-4265)
Initiated
2011-05-13
Root cause
Nonconforming Material/Component
Terminated
2012-07-13
Distribution
Worldwide distribution: (USA) Nationwide including the states of PA. NJ, NY, MA, VA, TX, KS, FL, OH, SC, MN, CA, TN and NV; and the countries of Canada, Colombia, France, Italy, Japan, Netherlands and United Kingdom.
Recall number
Z-2778-2011
Recall status
Terminated
CDRH record ID
100743
FDA classified
2011-07-11
Recall event number
58970