Open original FDA record ↗Recall record
- FDA identifier
- Z-2778-2020
- Company in FDA record
- Anjon Holdings
- FDA record creation date
- Aug 10, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
High rate of galling of one threaded component.
Source facts
- Action
- The one unit released into the United States has already been returned, unused, to Anjon. No uncorrected product remains in distribution or in use in the USA. No units subject to the failure mode have been released into distribution within Japan. All units shipped to Japan are being either returned to Anjon for reprocessing to remove the failure mode; or are to corrected by the addition of the instruction to use sterile surgical lubricant on the thread of the Slider. JMDM is the license holder for the product within Japan and will not release the product until resolution to its satisfaction of the issue.
- Initiated
- 2020-06-18
- Root cause
- Device Design
- Terminated
- 2020-11-25
- Distribution
- Worldwide distribution - US Nationwide distribution including in the state of MI and the country of Japan for training purposes.
- Recall number
- Z-2778-2020
- Recall status
- Terminated
- CDRH record ID
- 182271
- FDA record created
- 2020-08-10
- Recall event number
- 85958