Z-2778-2020: Anjon Bremer Molded Crown. bone fixation fastener for spinal use. REF number 100048 - Product Usage: Smo

Recall record Z-2778-2020 for Anjon Holdings. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2778-2020
Company in FDA record
Anjon Holdings
FDA record creation date
Aug 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

High rate of galling of one threaded component.

Source facts

Action
The one unit released into the United States has already been returned, unused, to Anjon. No uncorrected product remains in distribution or in use in the USA. No units subject to the failure mode have been released into distribution within Japan. All units shipped to Japan are being either returned to Anjon for reprocessing to remove the failure mode; or are to corrected by the addition of the instruction to use sterile surgical lubricant on the thread of the Slider. JMDM is the license holder for the product within Japan and will not release the product until resolution to its satisfaction of the issue.
Initiated
2020-06-18
Root cause
Device Design
Terminated
2020-11-25
Distribution
Worldwide distribution - US Nationwide distribution including in the state of MI and the country of Japan for training purposes.
Recall number
Z-2778-2020
Recall status
Terminated
CDRH record ID
182271
FDA record created
2020-08-10
Recall event number
85958

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗