Skip to content

Public FDA record

Z-2778-2026: MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)

Recall record Z-2778-2026 for Medtronic MiniMed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2778-2026
Company in FDA record
Medtronic MiniMed, Inc.
FDA classification date
Jul 23, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

Source facts

Action
On 6/16/2026 In-App Push Notifications were sent to customers entitled, Important Field Safety Corrective Action required on your MiniMed Go app. informing them a software upgrade is available to ensure important nighttime notifications are received. Correction notices were also emailed to customers informing them to update to version 1.2.0 or higher. For iOS Users: To confirm the latest app version on your iOS device, follow the steps below: 1. Open the App Store. 2. Tap your profile icon at the top of the screen. 3. Scroll to see pending updates. 4. Tap Update next to the MiniMed Go app icon. For Android Users: To confirm the latest app version on your Android device, follow the steps below: 1. Open the Google Play Store. 2. Tap Menu, then select Manage apps & device. 3. Apps with an update available are labeled "Update available." You can also search for the MiniMed Go app. 4. Tap Update next to the MiniMed Go app icon. Acknowledge that you have read and understood the notification by clicking the acknowledgement button or visiting https://www.minimed.com/en-us/safety-notices/notice31-letter If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1.
Initiated
2026-06-16
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states and territories of PR, OH, WA, IL, NC, PA, VA, MN, ID, CA, UT, MI, NJ, MS, SC, CO, AZ, OR, AR, MO, TX, AL, FL, NY, IN, TN, IA, WI, MA, LA, OK, GA, KS, NM, CT, KY, NH, NV, MD and the countries of Australia, Switzerland, Germany, Denmark, Spain, Finland, Iceland, Norway, Sweden.
Recall number
Z-2778-2026
Recall status
Open, Classified
CDRH record ID
221103
FDA classified
2026-07-23
Recall event number
99228

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗