Open original FDA record ↗Recall record
- FDA identifier
- Z-2781-2015
- Company in FDA record
- Salter Labs
- FDA classification date
- Sep 17, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Inspiratory valve cap on the nebulizer reservoir tower may be loose and may detach before or during treatment.
Source facts
- Action
- Salter Labs sent a Medical Device Recall letter dated July 24 2015. The letter identified the affected product, problem and actions to be taken. The letter states that all affected product can be returned to Salter Labs for replacement at no charge with product specifically inspected for a tighter fit on the inspiratory valve cap. Customers were instructed to complete and return the Return Response Form as soon as possible. For any questions contact Salter Labs Customer Support at 800-421-0024 or the local Salter Labs representative.
- Initiated
- 2015-08-04
- Root cause
- Component change control
- Terminated
- 2016-05-06
- Distribution
- Worldwide Distribution - US Nationwide and in the countries of Greece, UK, Belgium, Australia, and Switerzland.
- Recall number
- Z-2781-2015
- Recall status
- Terminated
- CDRH record ID
- 139320
- FDA classified
- 2015-09-17
- Recall event number
- 71909