Open original FDA record ↗Recall record
- FDA identifier
- Z-2786-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jul 11, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Firm has discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation.
Source facts
- Action
- Siemens initated their recall of the device on March 31, 2001, by sending a Customer Safety Advisory Notice/Letter to affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that detectors will be installed in appropriate locations on the system, which will switch off the system's power electronics in the event of a malfunction in the power supply of the gradient coils, preventing damage to the system and its environment.
- Initiated
- 2011-03-31
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-01-08
- Distribution
- Nationwide Distribution including AL, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, and WI and Puerto Rico.
- Recall number
- Z-2786-2011
- Recall status
- Terminated
- CDRH record ID
- 100474
- FDA classified
- 2011-07-11
- Recall event number
- 58885