The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).
Source facts
Action
The Regional Business Director notified the customer by telephone and informed them that there was an issue with the newly released calcaneus plates. The 6 units were requested to be returned to Flower Orthopedics.
Initiated
2020-07-08
Root cause
Mixed-up of materials/components
Terminated
2020-10-08
Distribution
US Nationwide distribution including in the states of AZ, FL, and OH.
Recall number
Z-2787-2020
Recall status
Terminated
CDRH record ID
182524
FDA classified
2020-08-12
Recall event number
86023
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.