Skip to content

Public FDA record

Z-2787-2026: Medline convenience kits containing Bard Silastic Foley Catheters (French sizes 16, 18, 20) 5cc balloons.

Recall record Z-2787-2026 for Medline Industries, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2787-2026
Company in FDA record
Medline Industries, LP
FDA classification date
Jul 23, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.

Source facts

Action
On June 8, 2026, the firm began notifying customers through First-Class Mail and email. Customers were instructed to over-label affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
Initiated
2026-06-08
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of CA, FL, MD. PA, and TX.
Recall number
Z-2787-2026
Recall status
Open, Classified
CDRH record ID
221243
FDA classified
2026-07-23
Recall event number
99301

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.