A software anomaly has been identified with the Eclipse Monte Carlo (eMC) algorithm where not-equidistant data points in the open beam profile used during configuration may lead to inaccurate results.
Source facts
Action
Varian sent an URGENT MEDICAL DEVICE CORRECTION letter via certified mail dated May 20, 2011, to all affected customers.
***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remove the product from use and advise users to manually initiate the configuration process for resampling to be applied when a previous calculation model is copied. Customers were asked to sign and return the notice to confirm receipt.
For questions call USA and Canada 1-888-827-4265
For Europe call +41 41 749 8844
Initiated
2011-05-20
Root cause
Software design
Terminated
2012-07-13
Distribution
Worldwide Distribution - USA (nationwide)
Recall number
Z-2788-2011
Recall status
Terminated
CDRH record ID
100818
FDA classified
2011-07-11
Recall event number
59006