Open original FDA record ↗Recall record
- FDA identifier
- Z-2788-2018
- Company in FDA record
- Guangzhou Wondfo Biotech Co., Ltd.
- FDA record creation date
- Aug 12, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
GUANGZHOU WONDFO BIOTECH discovered on the TCube One Step Multi-Drug Oral Fluid Test Cube the potential for failure of the saliva collection device. The use of the saliva collection device could cause the foam tip to become dislodged in the user's mouth creating a choking hazard.
Source facts
- Action
- The firm notified its customers on or about 05/22/2018 via a letter entitled "URGENT MEDICAL DEVICE RECALL". The firm advised the end user to examine their inventory and quarantine product subject to recall. In addition, if they have further distributed this product, please identify the customers and notify them at once of this product recall. The notification to customers may be enhanced by including a copy of the recall notification letter. The firm advised the end user to follow the directions indicated in the "Customer Required Action" section below and return the completed response form as soon as possible by faxing. to 123-456-7890,or by scanning and e-mailing to TOX.FCA@alere.com. They also stated to complete and return the form even if the customer may have consumed or distributed all affected product. Any questions should be directed toward Alere Toxicology Services Phone: 123-456-7890, Email:FCA.TOX@alere.com.
- Initiated
- 2018-05-18
- Root cause
- Nonconforming Material/Component
- Distribution
- US Distribution: in the state of VA.
- Recall number
- Z-2788-2018
- Recall status
- Open, Classified
- CDRH record ID
- 165501
- FDA record created
- 2018-08-12
- Recall event number
- 80443