Open original FDA record ↗Recall record
- FDA identifier
- Z-2791-2011
- Company in FDA record
- Abbott Vascular
- FDA classification date
- Jul 12, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Five lots of Guide Catheters and Accessory products were processed through Ethylene Oxide (ETO) sterilization with additional plastic material covering the pallet, which is not consistent with validated packaging configuration for ETO sterilization.
Source facts
- Action
- Abbott Vascular sent an Urgent Device Recall letter, dated January 7, 2011, noting the products being recalled and the reasons for recall. The firm also provided the following reconciliation instructions to its territory manager: -Help customer identify aflected product. - CaIl Abbott Vascular Customer Service (800) 221-9902 to receive RGA number. Record RGA number below. - Fax this completed form to (951) 914-3826. - Return a copy of this completed form with the returned product
- Initiated
- 2011-01-07
- Root cause
- Process control
- Terminated
- 2012-06-26
- Distribution
- Nationwide distribution, and Ireland.
- Recall number
- Z-2791-2011
- Recall status
- Terminated
- CDRH record ID
- 100790
- FDA classified
- 2011-07-12
- Recall event number
- 58979