Skip to content

Public FDA record

Z-2791-2011: Abbott Vascular ACS .035 Torque Device Part Number : 1003279. For use in cardiac diagnostic or thera

Recall record Z-2791-2011 for Abbott Vascular. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2791-2011
Company in FDA record
Abbott Vascular
FDA classification date
Jul 12, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Five lots of Guide Catheters and Accessory products were processed through Ethylene Oxide (ETO) sterilization with additional plastic material covering the pallet, which is not consistent with validated packaging configuration for ETO sterilization.

Source facts

Action
Abbott Vascular sent an Urgent Device Recall letter, dated January 7, 2011, noting the products being recalled and the reasons for recall. The firm also provided the following reconciliation instructions to its territory manager: -Help customer identify aflected product. - CaIl Abbott Vascular Customer Service (800) 221-9902 to receive RGA number. Record RGA number below. - Fax this completed form to (951) 914-3826. - Return a copy of this completed form with the returned product
Initiated
2011-01-07
Root cause
Process control
Terminated
2012-06-26
Distribution
Nationwide distribution, and Ireland.
Recall number
Z-2791-2011
Recall status
Terminated
CDRH record ID
100790
FDA classified
2011-07-12
Recall event number
58979

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗