Open original FDA record ↗Recall record
- FDA identifier
- Z-2791-2018
- Company in FDA record
- Cook Inc.
- FDA record creation date
- Aug 12, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
This lot of VentriClear failed endotoxin testing. Potential adverse events include immune responses ranging from non-specific febrile reaction to life-threatening systemic inflammatory response syndrome (SIRS).
Source facts
- Action
- On May 31, 2018, the firm contacted the single customer (a distributor) who received the entirety of the affected lot, via email. The email alerted the customer that the lot failed endotoxin testing and instructed the customer to quarantine the lot immediately. The firm then conducted a teleconference with the distributor to further discuss interim controls for future lot release. The recalling firm requested return of the affected lot. The distributor returned the lot to the firm. The recalling firm stated that none of the product was distributed to end users.
- Initiated
- 2018-05-31
- Root cause
- Release of Material/Component prior to receiving test results
- Terminated
- 2019-04-11
- Distribution
- One distributor in Indiana; product was not further distributed to end users.
- Recall number
- Z-2791-2018
- Recall status
- Terminated
- CDRH record ID
- 165597
- FDA record created
- 2018-08-12
- Recall event number
- 80492