Skip to content

Public FDA record

Z-2791-2018: VentriClear Ventricular Drainage Catheter Set, Cat. No. N-VVDC-01-ABRM Product Usage: The VentriClear Ve

Recall record Z-2791-2018 for Cook Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2791-2018
Company in FDA record
Cook Inc.
FDA record creation date
Aug 12, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This lot of VentriClear failed endotoxin testing. Potential adverse events include immune responses ranging from non-specific febrile reaction to life-threatening systemic inflammatory response syndrome (SIRS).

Source facts

Action
On May 31, 2018, the firm contacted the single customer (a distributor) who received the entirety of the affected lot, via email. The email alerted the customer that the lot failed endotoxin testing and instructed the customer to quarantine the lot immediately. The firm then conducted a teleconference with the distributor to further discuss interim controls for future lot release. The recalling firm requested return of the affected lot. The distributor returned the lot to the firm. The recalling firm stated that none of the product was distributed to end users.
Initiated
2018-05-31
Root cause
Release of Material/Component prior to receiving test results
Terminated
2019-04-11
Distribution
One distributor in Indiana; product was not further distributed to end users.
Recall number
Z-2791-2018
Recall status
Terminated
CDRH record ID
165597
FDA record created
2018-08-12
Recall event number
80492

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗