If two devices are plugged into the same channel/port in the generator, both devices activate. If one device is left laying on the patient, the result could be burning of the patient.
Source facts
Action
The firm sent letters in June 2018, informing their customers of the issue. The firm will send labels that can be applied to the product to remind them not to plug 2 active devices into the same channel at a time.
Updated 08/13/19
Starting on 06/29/18, updated Medical Device Correction Notifications and labels were mailed via second day service. Customers were asked to complete and return Business Reply Forms.
Initiated
2018-05-15
Root cause
Use error
Terminated
2024-01-30
Distribution
Worldwide distribution. US Nationwide, Mexico, Europe, Middle East, Latin America, Africa, and Asia Pacific Markets
Recall number
Z-2792-2018
Recall status
Terminated
CDRH record ID
165204
FDA classified
2018-08-13
Recall event number
80327