Open original FDA record ↗Recall record
- FDA identifier
- Z-2793-2015
- Company in FDA record
- Ecolab Inc
- FDA classification date
- Sep 18, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The sterile packaging may contain small (less than 250 microns wide) channels in the seal of the pouch. If channels are present, the package integrity is compromised and there is a remote chance that a non-sterile drape may be used during a surgical procedure.
Source facts
- Action
- Consignees were sent on 8/18/2015 an Ecolab "Urgent Medical Device Recall" letter dated August 18, 2015. The letter was addressed to Recall Coordinator or Risk Manager. The letter described the problem and the product involved in the recall. Advised consignees to quarantine all affected lots, and notify their customers of the recall which is to be carried out to users level. Requested consignees to complete and return the enclosed Recall Response Form. For questions they can contact Customer Service at 1-800-824-3027 or e-mail customerservice@microtekmed.com.
- Initiated
- 2015-08-18
- Root cause
- Package design/selection
- Terminated
- 2017-08-09
- Distribution
- Worldwide Distribution -- US, including DC and PR; and, the countries of SAUDI ARABIA, LEBANON, GERMANY, HONG KONG, SINGAPORE, CANADA, and JAPAN.
- Recall number
- Z-2793-2015
- Recall status
- Terminated
- CDRH record ID
- 139446
- FDA classified
- 2015-09-18
- Recall event number
- 71957