Skip to content

Public FDA record

Z-2793-2018: NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A) Product Usage: The NxStage

Recall record Z-2793-2018 for NxStage Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2793-2018
Company in FDA record
NxStage Medical, Inc.
FDA record creation date
Aug 14, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is plugged in. The firm has received 7 reports of smoke, sparks, or flames coming from the Express Fluid Warmer since 2016. No adverse events have been reported.

Source facts

Action
On June 11, 2018, the firm distributed Medical Device Product Safety Notices to affected customers and dialysis centers. Patients were contacted either via email or UPS second day air. Dialysis centers were contacted via email. The recall communication alerted customers that the firm has received 7 reports of smoke, sparks, or flames coming from the Express Fluid Warmer since 2016. No adverse health consequences or medical interventions have been reported. The firm informed customers that they were updating the User Guide to add the following new warning language: "Warning: Do not use any fluid bags that are leaking, and use caution not to spill fluids near the warmer. If the warmer is plugged in, leaking or spilled fluids in the warmer power receptacle can cause an electrical short even if the warmer is not being used. This can lead to an electrical fire and risk of serious injury or death. To minimize this risk, do the following: " When not used during treatment, do not plug the Express Fluid Warmer into the cycler. " Check all fluid bags for leaks prior to and during use. " If fluid spills onto the warmer, unplug the warmer power cord from the rear power receptacle and contact Technical Support." Customers were asked to do the following: 1. Please follow the instructions on the attached reply form to confirm that you received this notice. 2. Please observe the above warning and unplug your Express Fluid Warmer when not in use. If you have any questions or comments, please feel free to contact Customer Service at 1-800-NXSTAGE.
Initiated
2018-06-11
Root cause
Use error
Terminated
2020-04-28
Distribution
Worldwide Distribution - US Nationwide in the countries of Australia, Canada, France, Germany, Italy, Kuwait, Netherlands, Saudi Arabia, Spain, Sweden, and UK.
Recall number
Z-2793-2018
Recall status
Terminated
CDRH record ID
165648
FDA record created
2018-08-14
Recall event number
80512

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗