Z-2793-2020: BRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 0405648100

Recall record Z-2793-2020 for Brainlab AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2793-2020
Company in FDA record
Brainlab AG
FDA record creation date
Aug 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect manufacturer calibration

Source facts

Action
1. Refer to the appendix for the list of Brainlab Curve or Kick Navigation Systems with an affected camera installed to determine if your navigation system is affected. 2. For Cranial or ENT navigation software in combination with an affected system camera installed: Do not use the Z-touch Laser Pointer for patient registration. Use other instruments with the standard instrument pivoting method for Surface Matching or Landmark Registration, or Automatic Image Registration (AIR) if available. 3. Please continue to follow the instructions and warnings as described in the user guide. Especially relevant is the following warning, in addition to the instructions on the registration verification page appearing in the Cranial or ENT software: Verify accuracy at multiple anatomical landmarks, especially in the region of interest, as it may differ from the accuracy verified on the skin surface. If the region of interest is not accessible, verify in areas as close as possible to the region of interest. 4. In order to avoid nuisance with Spine & Trauma 3D or Orthopedic Navigation software in combination with an affected system camera installed: Do not attempt to use the optional Disposable Clip-On Remote Control. For patient registration, acquire registration points with the standard instrument pivoting method; or for Spine & Trauma 3D use Automatic Image Registration (AIR) if available.
Initiated
2020-07-16
Root cause
Process control
Terminated
2021-07-27
Distribution
USA: CA, TX, RI OUS: Armenia; Australia; Brazil; Bulgaria; Canada; Finland; France; Germany; Hong Kong; India; Israel; Italy; Japan; Mongolia; Morocco; Spain
Recall number
Z-2793-2020
Recall status
Terminated
CDRH record ID
182574
FDA record created
2020-08-13
Recall event number
86046

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗