Z-2794-2020: Philips HeartStart MRx, Monitor/Defibrillators with model numbers M3535A, M3536A, M3536M, M3536MC, M3536M

Recall record Z-2794-2020 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2794-2020
Company in FDA record
Philips North America, LLC
FDA record creation date
Aug 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The therapy switch may fail, resulting in the device exhibiting the following behaviors: the device may not perform the selected function, the therapy knob may not change to the energy setting selected, or the device may deliver a shock with an energy level different from the setting selected by the user. These device behaviors could result in a delay in therapy or a failure to deliver the intended therapy.

Source facts

Action
On July 16, 2020, the firm distributed an Urgent Medical Device Correction letter to all affected customers. Customers were informed of the product issue and were instructed to do the following: - The MRx is safe to use and can remain in service if the device does not exhibit any of the behaviors described in this Notice (the device may not perform the selected function, the therapy knob may not change to the energy setting selected, or the device may deliver a shock with an energy level different from the setting selected by the user.) - If an MRx device experiences a switch failure as described, contact your local Philips representative, who can arrange for a replacement of the switch at no charge. - Continue to perform the Shift and Operational Checks described in the Instructions for Use, Maintenance Chapter - To detect possible switch failures as early as possible, conduct Weekly Shock Tests (part of the Shift Check in the IFU) on a daily basis - It is not necessary to perform a Shock Test on days when an Operational Check is performed, since an Operational Check includes a Shock Test. - Keep a copy of the customer letter with each HeartStart MRx Instructions for Use. - To acknowledge receipt of this notification, please complete and fax the Customer Reply Form to the firm. If the MRx Therapy Switch fails, Philips will arrange for the replacement of the switch at no charge to the customer.
Initiated
2020-07-16
Root cause
Process control
Terminated
2021-08-03
Distribution
Worldwide distribution.
Recall number
Z-2794-2020
Recall status
Terminated
CDRH record ID
182101
FDA record created
2020-08-13
Recall event number
85743

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗